| Model Number |
834F75 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during use of the swan ganz catheter, it provided high cvp values.The catheter was placed in the or shortly before the inaccurate values were seen in the unit.After re-zeroing and troubleshooting the cvp transducer, the cvp line was moved to the double lumen central venous catheter (cvc) and the reading there was an expected value.There were microcracks in the blue lumen at the connection with the proximal injectate lumen.The rn was still able to check thermodilution utilizing the blue pa line but monitored cvp on double lumen cvc instead.There was no allegation of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional fda product codes include:
dqe- catheter, oximeter, fiberoptic.Dqo- catheter, intravascular, diagnostic.Kra- catheter, continuous flush.The device evaluation is anticipated.A supplemental report will be forthcoming when the investigation is completed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|
| |
|