• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Ethicon Endo-Surgery, LLC. ECHELON; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Ethicon Endo-Surgery, LLC. ECHELON; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number ECH45S
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a vats lobectomy, the surgeon experienced difficulty closing the device.A second device was also utilized in the procedure and difficulty closing was experienced.Both devices were loaded with a green reload which worked as intended.There was no patient consequence reported.No additional information provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 9/18/2026.D4: batch #a98n9e.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ech45s device was returned with the anvil bent upwards and without reload present.No functional test was performed due to the condition of the device.It is possible that the device was clamped over an excess of tissue causing the anvil to bend.The device event log was reviewed.The log indicates that the device had a high force to fire for one of the clinical firings, indicating that the device was firing over an obstruction, or too thick of tissue.In addition, the log indicates that articulation was attempted while the device was closed.Articulation function is blocked when the device is closed to prevent inadvertent patient injury.When the home button is pressed while the jaws are closed the device will attempt to retract the knife.When any of the other articulation buttons are pressed while the jaws are closed the device will provide haptic feedback indicating that articulation function is blocked.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number a98n9e, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished #ech45s, with lot/batch #a98r20, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR  00969
Manufacturer (Section G)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 00969
PR   00969
+1(303)552-6892
MDR Report Key50820372
Report Number3005075853-2026-07695
Device Sequence Number10648101
Product Code GAG
UDI-Device Identifier10705036035008
UDI-Public(01)10705036035008(17)280831(10)A98R20
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K213633
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberECH45S
Device Lot NumberA98R20
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/28/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2026
Initial Report FDA Received Date09/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/02/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-