| Model Number |
P7800A |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Baxter received a report that the compella cap us scale pd ppm failed to stop.This occurred during baxter servicing/testing.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The baxter technician found the power drive needed to be replaced.Per the hillrom service manual the compella bed requires an effective maintenance program.We recommend that you perform annual preventative maintenance.Pay particular attention to safety features, which include but are not limited to: controls return to off or the neutral position when released.Make sure that the controls are in good condition and that the membranes are not worn or torn.Do the function checks.Make sure that the bed does not fail any of the function checks.Replace or repair parts as necessary.A search of the baxter maintenance records showed baxter performed preventative maintenance on this bed in aug 2025.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the intellidrive transport handles to resolve the reported event.Based on this information, no further action is required.
|
| |
|