MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION CORPORATION Linear? ST; Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION CORPORATION Linear? ST; Stimulator, Spinal-Cord, Totally Implanted For Pain Relief Back to Search Results
Model Number SC-2218-70
Medical Device Problem Codes High impedance (1291); Unexpected Therapeutic Results (1631)
Health Effect - Clinical Code Pain (1994)
Date of Event 05/22/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient had device interrogation and revealed failed contacts.Due to high impedances, the patient's areas of pain could not be stimulated.Consequently, the patient underwent a spinal cord stimulator (scs) replacement procedure.
 
Additional Manufacturer Narrative
Block b3: exact date unknown, event occurred three weeks ago from the appointment date 29may2026.Additional suspect medical device components involved in the event: model number/catalog number: sc-2218-70, serial number: (b)(6), batch/lot number: 7102549, model/catalog description: linear st lead kit 70 cm, unique identifier (udi): (b)(4).Model number/catalog number: sc-1216, serial number: (b)(6), batch/lot number: 765194, model/catalog description: wavewriter alpha 16 ipg kit, unique identifier (udi): (b)(4).Sc-2218-70 sn: (b)(6); sc-1216 sn: (b)(6).Investigation results: the devices were not returned for analysis; therefore, a technical analysis could not be performed.Device history record: it was confirmed these devices met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion: based on a thorough review of the reported complaint, an investigation conclusion code of cause not established was assigned to the investigation for all components.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Linear? ST
Common Device Name
Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
Boston Scientific Corporation
road 698, lot no. 12
dorado PR 00646 -260
PR   00646-2602
Manufacturer Contact
ricardo amaro castro
25155 rye canyon loop
valencia, CA 91355
+1(320)800-7570
MDR Report Key50820689
Report Number3006630150-2026-06888
Device Sequence Number10680893
Product Code LGW
UDI-Device Identifier08714729767749
UDI-Public(01)08714729767749(17)260821(10)7102620
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P030017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/21/2026
Device Model NumberSC-2218-70
Device Catalogue NumberSC-2218-70
Device Lot Number7102620
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2026
Initial Report FDA Received Date09/18/2026
Date Device Manufactured08/21/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
Patient SexMale
Patient Weight174 LB
-
-