| Model Number |
Patient Information Center iX |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Philips received a complaint on a patient information center ix indicating that the alarm tones are not audible on pic ix surveillance computer located in er.The device was in use at time of event, no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis was performed by the remote service engineer did some troubleshooting with the customer biomed on the pic ix surveillance workstation and was able to replicate the customer issue where the alarm tones were not audible.During the investigation it identified a faulty network cable connected to the remote external speaker.It was recommended to replace the cable.Based on the results of the information available, it was determined that the product has malfunctioned and the most likely cause of the reported problem was a defective network cable.The issue was resolved by replacing the faulty network cable associated with the remote external speaker.Replacement of the cable restored speaker operation and alarm audibility.Subsequent testing confirmed proper alarm annunciation and the system was returned to service meeting specifications.
|
| |
|