| Catalog Number |
08057443190 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable cholesterol gen.2 results from the cobas c 503 analytical unit.The initial result was 3.41 mg/dl, and the repeat result was 295 mg/dl.The questionable result was not reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
The cobas c 503 analytical unit serial number was (b)(6).The field service engineer found that the lamp was not fully extended.He fully extended the lamp mechanism to seat against the incubation disk.Abnormal aspiration alarms were present in the analyzer alarm trace.Therefore, a pre-analytical issue cannot be excluded.The investigation is ongoing.
|
| |
|