| Model Number |
9-PDAP-03-02-L |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported on (b)(6) 2026 a 3mm x 2mm amplatzer piccolo occluder was selected for implant.The patient had a ductus with a minimal diameter of 1.4mm and length of 7mm.Transthoracic echocardiogram (tte) and fluoroscopy were used during the procedure.Upon fully releasing the occluder from the delivery cable, the device embolized into the pulmonary artery.The patient experienced moderate tricuspid regurgitation (tr).The device was snared and removed from the patient, which resolved the tr.A 4mm x 2mm amplatzer piccolo occluder was implanted.The patient status was stable.The physician believed the compliant duct and steep angle of the patent ductus arteriosus takeoff caused the embolization.
|
| |
|
Additional Manufacturer Narrative
|
|
It is unknown if the device is returning for analysis.A follow-up report will be submitted with all additional relevant information.
|
| |
|