The reported issue was "high pressure , makes the medication pumps alarm for pressure changes." a sample was received for investigation, and investigation included review of recent complaint trends.The complaint was not confirmed; it was reported that the customer had utilized the administration set with an iv pump, and as these sets are not pressure-rated this could be the root cause of the issue.No further action was taken because objectionable trends were not seen for this issue, and division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|