MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; IV EXT SET,60",2.0ML, M/F LUERLOCK,CLAMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; IV EXT SET,60",2.0ML, M/F LUERLOCK,CLAMP Back to Search Results
Catalog Number DYNDTN1603
Medical Device Problem Code Increase in Pressure (1491)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/27/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "high pressure , makes the medication pumps alarm for pressure changes".
 
Additional Manufacturer Narrative
The reported issue was "high pressure , makes the medication pumps alarm for pressure changes." a sample was received for investigation, and investigation included review of recent complaint trends.The complaint was not confirmed; it was reported that the customer had utilized the administration set with an iv pump, and as these sets are not pressure-rated this could be the root cause of the issue.No further action was taken because objectionable trends were not seen for this issue, and division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
IV EXT SET,60",2.0ML, M/F LUERLOCK,CLAMP
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
nicholas kourkine
three lakes drive
northfield, IL 60093--275
MDR Report Key50823536
Report Number1417592-2026-45175
Device Sequence Number21510974
Product Code FPA
UDI-Device Identifier20080196415073
UDI-Public20080196415073
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K093773
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYNDTN1603
Device Lot Number(10) 0332210A08
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/31/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/27/2023
Initial Report FDA Received Date09/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-