(b)(4)
date sent: 9/18/2026
b3: only event month and year known: september 2026
d4: batch #unk
d4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.H4: the device manufacture date is currently not available.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent:
was there a change in the procedure? if yes, please explain.Did the damage occur right out of the packaging, during loading/unloading, or in the operative field?
what part of the device was broken?
did the damage occur right out of the packaging or in the operative field?
did any pieces fall into the patient?
if yes, were they retrieved?
will all pieces be returned with the device?
if no, were they discarded?
attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
(b)(4).Date sent: 9/28/2026.D4: batch #919d96.Additional information: d4 (expiration date), h4.Correction: d4 (udi).Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one gst45d reload (d) was received fully fired with damaged on the edge from left side.No functional test was performed due the condition of the reload.It is possible that the damage noted on the returned reload was due to improper handling of the reload.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 919d96, and no non-conformances were identified.
|