| Catalog Number |
TRG500L |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that "exam gloves are tearing when being donned.There is a visible yellow line that is the failure point.".
|
| |
|
Additional Manufacturer Narrative
|
|
The customer reported gloves that tore during donning, and the complaint transcript stated there was a visible yellow line that is the failure point.No sample nor image was available for confirmation or further review; the vendor's investigation identified no defects or abnormalities within batch records or retention samples, and trending did not indicate a trend of 'tearing during donning' or similar occurring in the field.Based on the results of the investigation, medline was unable to confirm the complaint issue nor that the product failed to meet specifications.A cause was not determined because a sample was required to determine cause, and no adverse trending was identified.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|
| |
|
Search Alerts/Recalls
|
|
|