The account reported that the wings of the 5ml syringe were breaking during use, and the most recent issue resulted in the provider cutting himself; the customer contact reported the finger was cut through the glove, approximately 0.25¿ laceration, with no impact to the patient or the procedure.A sample was received for investigation, and the 5 ml syringes were removed from three sealed kits and inspected for visible defects, but nothing was noted; force gauge testing found one side of the finger flange broke off on each of two sample syringes with an average of 17.9 lb of force, and production records and pack build review found no issues that would contribute to the reported issue.The reported problem was confirmed during sample testing.Based on the complaint details, the root cause is considered to be a vendor product issue relating to the component manufacturing process.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|