Customer reported: "glove ripped"; "was there injury? n0 - error occurred but did not reach the individual." the account was able to provide a sample of an already torn glove; no unused samples were provided for additional testing, and the glove was evaluated for fingertip, palm, and cuff thickness.Measurements for length and width could not be obtained because the glove was torn in half, and all measured thickness values were found to be within the applicable product specifications outlined in astm d6319.With the provided sample, medline confirmed the reported issue; however, medline was unable to determine the root cause, and the quality team will continue to monitor for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|