Model Number NVRO-25
Medical Device Problem Code Perivalvular Leak (1457)
Health Effect - Clinical Code Aortic Valve Insufficiency/ Regurgitation (4450)
Date of Event 08/21/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Clinical information: (b)(4) - envision ide study, patient site (b)(6).It was reported that on (b)(6) 2026, a 25mm navitor vision valve was chosen for implant using a small flexnav delivery system.A pre-implant balloon valvuloplasty was done with a 24mm non-abbott balloon.The final implant depth at non-coronary cusp (ncc) was 3.46mm and left coronary cusp (lcc) was 7.43mm.The patient was discharged from the hospital the following day.On (b)(6) 2026, the 30 days tte noted a moderate paravalvular leak.No treatment was required.
 
Additional Manufacturer Narrative
The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Navitor Vision
Common Device Name
Aortic Valve, Prosthesis, Percutaneously Delivered
Manufacturer (Section D)
Abbott Medical
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
St. Jude Medical #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key50828044
Report Number2135147-2026-06354
Device Sequence Number10680143
Product Code NPT
UDI-Device Identifier05415067045782
UDI-Public(01)05415067045782(17)290222
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P190023
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberNVRO-25
Device Catalogue NumberNVRO-25
Device Lot Number11329942
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2026
Initial Report FDA Received Date09/18/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/23/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-