| Model Number |
K-LESS T-ROPE W/DRV, SYN REPR, TI |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Additional information was received on 21-sep-2026:\
the tightrope xp implant system was fully implanted before the issue was identified.Following deployment, the button position was confirmed after placement.During adjustment of the button, it penetrated the second cortical bone.No damage, deformation, breakage, fraying, or other visible abnormalities were observed on the device components.\
\
the affected implant was completely removed, and no device fragments were generated or retained in the patient.The procedure was prolonged by approximately one hour due to the event, resulting in approximately one additional hour of anesthesia administration.Alternative fixation was achieved using hardware from another manufacturer.\
\
the patient's bone quality was reported as good.The hospital stay was not extended, and no additional treatment, monitoring, or follow-up related to the event has been reported.No patient injury or adverse outcome was reported.Information regarding the patient's postoperative status and outcome is not available.\
\
additional information was received on 24-sep-2026:\
this was discovered during an ankle and syndesmosis arthroscopy procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3, h3, h6\
\
investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|