| Model Number |
TSGC0801 |
| Medical Device Problem Codes |
Leak/Splash (1354); Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported on (b)(6) 2026 that the patient presented with grade 3+ functional tricuspid regurgitation (tr) for a triclip procedure.During the procedure, after removal of dilator loss of column and air were observed in hemostasis valve of triclip steerable guide catheter(tsgc).The tsgc was retracted into atrium.Aspiration was performed for removal of air.The loss of column was observed again.The tsgc was replaced with another device to complete the procedure.The tr was reduced to grade 1 post procedure.There was no clinically significant delay reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|