Model Number RT380 |
Device Problem
Ventilation Problem in Device Environment (3027)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The complaint device was recently returned to fisher & paykel healthcare in (b)(4) for evaluation.We will provide a follow up report upon completion of our investigation.
|
|
Event Description
|
A hospital in the (b)(4) reported that a patient on an rt380 adult dual-heated evaqua2 breathing circuit was unable to be ventilated.The hospital further reported that the patient was manually ventilated while they replaced the breathing circuit.No patient consequence was reported.
|
|
Manufacturer Narrative
|
(b)(4).Method: the complaint rt380 adult dual heated evaqua2 breathing circuit was returned to fisher & paykel healthcare in (b)(4) where it was visually inspected and pressure tested.Results: visual inspection of the returned circuit revealed that the filter material was discoloured and occluded.The pressure test revealed that the breathing circuit was within specification.A lot check could not be carried out as the lot information was not provided.Conclusion: based on the inspection carried out, the problem reported by the customer was most likely due to the use of an occluded filter in the set up.It is likely that the discolouration and occlusion of the filter material was the result of nebulized drugs.All rt380 breathing circuits are visually inspected and pressure tested for leaks before releasing for distribution.Any breathing circuit which fails any of these tests is discarded.This suggests the damage noted on the filter was caused after the circuit was released for distribution.The user instructions supplied with the rt380 breathing circuit state: "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "set appropriate ventilator alarms." "change filter if noticeable deterioration occurs, following standard hospital procedure." "when nebulized drugs are used resistance to flow should be monitored and the filter replaced, following standard hospital procedure.".
|
|
Event Description
|
A hospital in the (b)(6) reported that a patient on an rt380 adult dual-heated evaqua2 breathing circuit was unable to be ventilated.The hospital further reported that the patient was manually ventilated while they replaced the breathing circuit.No patient consequence was reported.
|
|
Search Alerts/Recalls
|