| Model Number |
M00547010 |
| Medical Device Problem Codes |
Burst Container or Vessel (1074); Inflation Problem (1310); Use of Device Problem (1670); Failure to Deflate (4060)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an extractor pro rx retrieval balloon was attempted to be used during a cholangiogram procedure to treat gall stones performed on (b)(6) 2026.During the procedure, it was reported that the physician and nurses were unable to deflate device.It was reported that they overfilled in order to pop balloon to extract it.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event and patient's condition at the conclusion of the procedure was reported to be stable.Note: it was reported that, they overfilled in order to pop balloon to extract it.However, per the instruction for use (ifu): inflate the balloon to the appropriate size indicated for the device.Do not exceed the maximum balloon diameter as indicated on the enclosed pre-staked syringe when inflated.
|
| |
|
Additional Manufacturer Narrative
|
|
Block d4, h4:
detailed product information was not provided to bsc.We completed a good faith effort to obtain the information.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other product specific information.If additional details become available, a supplemental report will be submitted.Block h6:
imdrf device code a140101 captures the reportable event of balloon failure to deflate.Imdrf device code a1406 captures the reportable event of balloon overinflated.
|
| |
|