|
The customer reported the issue as ¿not tightening,¿ with blood ¿dripping occasionally¿ and ¿something is seems off on the connection port.¿ at this time, a sample(s) was not received for investigation; investigation included review of recent complaint trends, and without the sample or photo, it is difficult to truly identify the root use of the issue.The complaint was not confirmed, and ¿sample required to determine cause.¿ no further action was taken because objectionable trends were not seen for this issue, and division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|