| Model Number |
1847ATTSS |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that prior to use, the attachment rejects the burr and spits out oil.It was reported that there was no patient or staff impact.Additional information was received stating that loose distal bearings were found during evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis :evaluation determined that customer allegation of oil leakage and reject burr is confirmed.The likely cause was identified as due to loose bearing at distal.It was also noted that there was difficulty to insert testing burr.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|