| Catalog Number |
10445160 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that erroneously depressed calcium (ca) results on two patient samples were obtained on a dimension vista 1500 system.For sample id (b)(6), the initial result was not reported to the physician.This sample was repeated on an alternate dimension vista 1500 system.While the initial result of sample id (b)(6) was reported to the physician and not questioned.This sample was repeated on both original and on alternate dimension vista 1500 systems.All the repeat results obtained were higher, correlated clinically, and considered correct.The repeat results from alternate system were reported to the physician.There are no known reports of patient intervention or adverse health consequences due to the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
A united states (us) customer contacted the siemens customer care center (ccc) regarding erroneously depressed calcium (ca) results on two patient samples obtained on a dimension vista 1500 system.Siemens evaluated the instrument data and the information provided by the customer.The calibration results and quality control (qc) results were acceptable on the day of the event.The review of instrument data found no abnormalities with the instrument performance related to the reported event at the time of samples ran and the data indicated that the erroneous patient results originated from one of the well map position in flex sequence.No other discrepancies were observed in the remaining well positions within the same flex sequence or other flex sequences.The customer blew a new well, ran qc, and found it to be acceptable.Based on the available information, the probable cause of the event is consistent to the well position of the flex sequence.The reported event was resolved with standard troubleshooting actions.The customer is operational.The device is performing according to specifications.No further evaluation is required.
|
| |
|
Search Alerts/Recalls
|