| Catalog Number |
EBUS00001 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It is reported during endobronchial ultrasound case, endobronchial ultrasound scope (b)(6) broke related to thoracent/areus endobronchial ultrasound aspiration needle.Scope working at beginning of case.Passed needle through channel, scope still working.Attempted to deploy needle and image on screen became pixelated.Removed needle and attempted to reinsert/redeploy, and pixelation on screen became worse.Pixelation on screen became worse with every attempt to redeploy needle.
|
| |
|
Additional Manufacturer Narrative
|
|
After micro-tech received this incident, we reviewed the manufacturing documents from the device history record and found no abnormalities that would contribute to this issue.We are trying to get more informaiton for investigation.We will fill a follow-up report when this incident investigation is done.
|
| |
|