| Model Number |
0998-00-0800-52 |
| Medical Device Problem Codes |
Fire (1245); Failure to Power Up (1476); Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the customer that prior to use the cardiosave intra-aortic balloon pump (iabp) was repeatedly switching off and on intermittently; electrical burning smell was detected from the unit.There was no patient involvement reported.No harm was reported.A getinge field service engineer (fse) was dispatched to evaluate the issue.Fse found extensive internal electrical damage and all internal boards have blown.The device did not power on, displayed no lights or indications on the main board, and emitted a strong electrical burning smell.Fse was unable to get any information of the unit due to an internal burnout.The complaint involves electrical shutdown, possible fuse/wiring failure, and other electrical component damage.
|
| |
|
Additional Manufacturer Narrative
|
|
Event site postal code: (b)(6).Event site telephone: (b)(6).Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|