| Model Number |
120V/60 HZ |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Livanova deutschland received a report of a 3t heater/cooler which suddenly shut off during usage.The event occurred during a procedure.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11:
a livanova field service engineer (fse) was dispatched on site and the event was not confirmed.A full functional verification and electrical safety inspection was performed onto involved unit, and all measured parameters were found to be within manufacturer-specified operating tolerances and standard specifications.Consequently, unit was returned back to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|