| Model Number |
SC-2317-50 |
| Medical Device Problem Codes |
High impedance (1291); Charging Problem (2892)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/29/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient was unable to have a magnetic resonance imaging (mri) due to concern for abnormal lead impedance from the spinal cord stimulator (scs) device.It was also reported that the patients implantable pulse generator (ipg) was no longer keeping a charge.The patient underwent a revision procedure wherein the ipg, scs leads and clik anchor were replaced.The patient was reported to be doing well post operatively.The explanted device components were not returned due to facility policy.
|
| |
|
Additional Manufacturer Narrative
|
|
Block d.2b; additional applicable product code: qrb
additional suspect medical device components involved in the event:
model number/catalog number: sc-2317-50
serial number: (b)(6)
batch/lot number: 7077652
model/catalog description: infinion cx lead kit, 50cm
unique identifier (udi) #: (b)(4)
model number/catalog number: sc-1232
serial number: (b)(6)
batch/lot number: 545316
model/catalog description: wavewriter alpha 32 ipg kit
unique identifier (udi) #: (b)(4)
model number/catalog number: sc-4318
serial number: n/a
batch/lot number: 29661764
model/catalog description: clik x anchor sterile kit
unique identifier (udi) #: (b)(4)
sc-2317-50 sn: (b)(6)/ (b)(6)
investigation results:
the devices were not returned for analysis; therefore, a technical analysis could not be performed.Device history record
it was confirmed these devices met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Investigation conclusion
based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|
| |
|