Model Number 500001002502
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when using the bd pyxis¿ medstation¿ 4000 integrated main had a failed space that did not open.The customer stated that this malfunction caused a delay in dispensing medication to patients.There were no adverse events or injuries reported based on this event.
 
Additional Manufacturer Narrative
A review of the complaint history for sn 14937423 was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn 14937423 was performed from the date of manufacture, 16-jun-2017 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the drawer was not released when selected.A field service engineer (fse) adjusted the railings and tested the drawer operation properly.The repair was verified by testing the system using both the hardware test application (hta) and the dispensing application.The system functioned as intended after the field service engineer repaired the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD Pyxis¿ MedStation¿ 4000 Integrated Main
Common Device Name
Automated Dispensing Cabinet
Manufacturer (Section D)
CAREFUSION 303, INC.
3750 torrey view ct
san diego CA 92130
Manufacturer (Section G)
CareFusion 303, INC.
10020 pacific mesa blvd
san diego CA 92121 -438
Manufacturer Contact
brett wilko
3750 torrey view ct
san diego, CA 92130
+1(858)617-2000
MDR Report Key50836502
Report Number2016493-2026-64617
Device Sequence Number10270552
Product Code BRY
UDI-Device Identifier10885403512636
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number500001002502
Device Catalogue Number306
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2026
Initial Report FDA Received Date09/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/16/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-