It was reported that balloon rupture occurred.The 100% stenosed target lesion was located in the mildly tortuous and mildly calcified subclavian vein.A 12.0 x 40, 75cm mustang balloon catheter was advanced for dilation.During the procedure, the balloon vertically ruptured from the tip to the middle upon first inflation at 8 atmospheres for 30 seconds.The device was removed only using normal method without any problem, and the procedure was completed with another of the same device.There were no patient complications as a result of this event, and the patient was in good condition post-procedure.
|