On (b)(6) 2026, a patient underwent a computed tomography guided percutaneous transhepatic cholangiography drainage (ptcd) catheter placement procedure using a navarre universal drainage catheter.During the procedure, it was reported that the sure-lok thread completely disengaged from the catheter.The procedure was completed using another device.There was no reported patient injury.
|
|
H11: e1 - (b)(6).H.4.Device manufacture date: unknown.Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this lot.Investigation summary: although the physical device was not returned for evaluation, two photos were provided for review.The first photo shows the physician holding an implanted drainage catheter in one hand, and on the other hand, physician holding thread which was completely separated from hub.The second photo shows a clinician holding a drainage catheter in one hand and the detached thread in the other hand, with glue visible at the end of the separated thread.Therefore, the investigation is confirmed for the reported thread break and material separation issues.A definitive root cause could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.Section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|