| Catalog Number |
IPU05015013P |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the procedure, during the first balloon inflation at nominal pressure, the inpact 018 drug coated balloon incurred a pinhole rupture.The pressure was slowly increased up to 6 atmospheres.Calcification and mild tortuosity were reported.A 6fr 45cm non-medtronic sheath and a non-medtronic guidewire were used.Embolic protection was not used.No damage was noted to the packaging and no issues were noted when removing the device from the hoop/tray.The device was prepped as per the ifu.The balloon did not detach/fragment.The balloon did not pass through a previously deployed stent.No resistance was encountered during delivery.The procedure was completed using a non-medtronic balloon.No patient symptoms or complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|