| Model Number |
0998-00-0800-52 |
| Medical Device Problem Codes |
Communication or Transmission Problem (2896); Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that prior to use, the cardiosave intra-aortic balloon pump (iabp) alarmed and stopped working.There is no information about patient involvement.No harm is reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that critical error code 115 occurred.The fse recommend to replace executive processor board.
|
| |
|
Additional Manufacturer Narrative
|
|
Event site name: (b)(6)
event site postal code: 119074.Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|