It was reported that during transcatheter aortic valve implantation with the evfxplus-29 transcatheter aortic valve, a pre-implant balloon aortic valvuloplasty (bav) was performed for calcification.Upon implant of the valve, frame underexpansion was observed, and a post-implant bav was performed with rapid pacing.During the bav, the balloon reportedly interacted with the outflow crown and the valve dislodged.Implant depth changed from 3 mm at the non-coronary cusp and left coronary cusp before dislodgement to 0 mm at both cusps after dislodgement.No paravalvular leak, gradient, or central regurgitation was reported.Subsequently, a second medtronic valve was implanted.
|
|
D10.Continuation - concomitant medical devices in use during the event include medtronic product id: d-evolutfx-2329; product lot number: 0013439726; product type: 0195-heart valves; implant date: not-implantable; explant date: na non-medtronic product id: balloon valvuloplasty catheter; product lot number: unknown; implant date: na; explant date: na.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|