| Model Number |
WA50040A |
| Medical Device Problem Codes |
No Display/Image (1183); Communication or Transmission Problem (2896); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the rigid video laparoscope no image during inspection for use.During the initial evaluation of the device it was observed, e218 (scope error) and the fiber bonding in the outer tube area (distal end) was damaged.No patient involvement was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The final investigation has been completed.The failure reported by the customer and identified during the inspection have been confirmed.A definitive root cause could not be identified.Based on the results of the investigation, the video cable is broken and therefore the image is not displayed and scope error e218 appears.The most probable cause of the complaint was traced to component failure.Olympus will continue to monitor the field performance of this device and evaluate any additional information received through post-market surveillance activities.
|
| |
|