| Model Number |
H74939171100470 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon burst occurred.The 100% stenosed target lesion was located in the mildly tortuous and moderately calcified arteriovenous (av) shunt.A 10.0 x 40, 75cm mustang balloon catheter was advanced for dilation.During the procedure, the balloon burst at 12 atmospheres before inflating to the nominal burst pressure (nbp).The procedure was completed with another of same device.There were no patient complications as a result of this event, and the patient was stable post-procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
D2b: pro code (product code): fge, lit.G4: premarket / 510(k) #: k141521, k141597.
|
| |
|