| Model Number |
500004500001 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medbank tower aux, the user was unable to issue the medication.The drawer opened; however, the cubie did not open.A medication package was suspected to be obstructing the cubie lid from opening properly.The customer stated that this malfunction occurred while dispensing the medication.However, there were no adverse events or injuries reported based on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 03-apr-2023 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, a technical support specialist (tss) found in mql bin locations the item assigned to bin 11 01 with qoh=0.The tss guided the customer through performing a cycle count of the bin as part of the troubleshooting process.The tss confirmed with the customer that, after the medication obstruction was removed, the cubie lid opened successfully and normal functionality was restored.The system functioned as intended after the technical support specialist investigated the issue.
|
| |
|