A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 23-oct-2025 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the older lumidigm bio id (biometric identifier) hardware/software version, along with an incorrect med app auto-launch configuration and excessive database bloat, which resulted in bio id authentication failures, the station becoming stuck on the care fusion startup screen, and users being unable to access medications.A technical support specialist (tss) applied the designated lumidigm shim update, the med app auto-launch configuration was corrected, and database maintenance was performed to optimize performance.User authentication and station functionality were verified, and the customer confirmed that all issues were resolved before case closure.The system functioned as intended after the technical support specialist troubleshot the device.
|