| Model Number |
500001002002 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medstation¿ 4000, auxiliary drawer 2.2 duplicate address detected.The customer stated that the malfunction occurred when a user tried to issue medication to the patient and caused a delay to patient care.However, there were no adverse events or injuries reported based on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 01-jun-2011 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the tower auxiliary door 4 exhibited clicking sounds when prompted open.A field service engineer (fse) replaced and installed door latch to resolve the issue.The system functioned as intended after the field service engineer repaired the device.
|
| |
|