ICU MEDICAL DE MEXICO, S. De R.L. De C.V. 7" (18 cm) Appx 0.52 ml, Smallbore Trifuse Ext Set w/3 MicroClave® Clear, 3 Clam; Stopcock, I.V. Set
|
Back to Search Results |
|
| Catalog Number |
MC330027 |
| Medical Device Problem Code |
Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The event involved a 7" (18 cm) appx 0.52 ml, smallbore trifuse ext set w3 microclave clear, 3 clamps, luer lock.It was reported that during a lab blood draw from the patient's picc (peripherally inserted central catheter), the syringe filled with air instead of blood while the nurse was drawing back waste.The nurse observed that air was entering the line from the hub of the trifuse.The trifuse was immediately clamped and removed.No air reached the patient.The patient remained stable with no changes in vital signs or neurologic function, and no harm was reported.No issues were noted while priming the trifuse.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been requested for evaluation; however, it has not yet been received.
|
| |
|