| Model Number |
PU-681RA |
| Medical Device Problem Code |
Application Program Problem (2880)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The biomedical engineer (bme) reported that the central nurse's station (cns) displayed a windows blue screen with machine_check_exception error.The unit was powered off / on.It booted for approximately 20 seconds, began loading windows, then displayed a second blue screen with a clock_watchdog_timeout error.They also noted that the fan at the bottom/rear of the tower did not appear to be spinning.Symptoms appear consistent with a hardware issue.Tech support (ts) recommended sending the unit to the repair center for evaluation.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The biomedical engineer (bme) reported that the central nurse's station (cns) displayed a windows blue screen with machine_check_exception error.The unit was powered off / on.It booted for approximately 20 seconds, began loading windows, then displayed a second blue screen with a clock_watchdog_timeout error.They also noted that the fan at the bottom/rear of the tower did not appear to be spinning.Symptoms appear consistent with a hardware issue.Tech support (ts) recommended sending the unit to the repair center for evaluation.No patient harm was reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
|
| |
|