| Lot Number |
03439H26 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the moderately tortuous superficial femoral artery (sfa).A 6.0 x 100 mm 200 cm ranger drug-coated balloon catheter was advanced for dilation.During the procedure, the balloon ruptured upon first inflation at nominal atmospheres.The device was removed without any problem using the normal method, and the procedure was completed with another of the same device.There were no patient complications as a result of this event, and the patient's condition was satisfactory after the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of adverse event related to patient condition.
|
| |
|