| Model Number |
74001 |
| Medical Device Problem Code |
Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
Electric Shock (2554)
|
| Date of Event |
09/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The nalu sales representative (rep) reported that the patient said that he's having issues with his therapy disc (td), that the amplitude settings are changing and giving him shocks.No further information was provided.There was no reported patient harm or injury.
|
| |
|
Additional Manufacturer Narrative
|
|
After conducting a post-acquisition, retrospective review of nalu peripheral nerve stimulation complaint records, boston scientific determined that this event meets mdr reporting criteria.The following mdr information is being submitted as part of capa investigation pcar42 remediation activities.The therapy disc was returned for evaluation.The td passed ate and did not show signs of magnetization or increased rf power.Evaluation was unable to reproduce the reported amplitude and shock issue; therefore, the root cause of the reported event could not be determined.
|
| |
|