Nalu Medical, Inc Nalu Neurostimulation System; Peripheral Nerve Stimulator Back to Search Results
Model Number 74001
Medical Device Problem Code Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code Electric Shock (2554)
Date of Event 09/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
The nalu sales representative (rep) reported that the patient said that he's having issues with his therapy disc (td), that the amplitude settings are changing and giving him shocks.No further information was provided.There was no reported patient harm or injury.
 
Additional Manufacturer Narrative
After conducting a post-acquisition, retrospective review of nalu peripheral nerve stimulation complaint records, boston scientific determined that this event meets mdr reporting criteria.The following mdr information is being submitted as part of capa investigation pcar42 remediation activities.The therapy disc was returned for evaluation.The td passed ate and did not show signs of magnetization or increased rf power.Evaluation was unable to reproduce the reported amplitude and shock issue; therefore, the root cause of the reported event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Nalu Neurostimulation System
Common Device Name
Peripheral Nerve Stimulator
Manufacturer (Section D)
Nalu Medical, Inc
2320 faraday ave
suite 100
carlsbad CA 92008 -724
Manufacturer (Section G)
Nalu Medical, Inc
2320 faraday ave
suite 100
carlsbad CA 92008 -724
Manufacturer Contact
rebecca hess
2320 faraday ave
suite 100
carlsbad, CA 92008--724
MDR Report Key50839514
Report Number3015425075-2026-00698
Device Sequence Number21349637
Product Code GZF
UDI-Device Identifier00812537035444
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K183579
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number74001
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer12/06/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2022
Initial Report FDA Received Date09/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-