| Model Number |
BF-XP190 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the bronchovideoscope exhibited black images on the screen.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection and the reported failure was¿confirmed.The device has evidence of low angulation and no image/black out upon angulation.A root cause could not be identified.Based on the results of the investigation, it is likely component failure led to the malfunction.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|