Model Number G138
Medical Device Problem Codes Over-Sensing (1438); Pacing Problem (1439)
Health Effect - Clinical Codes Bradycardia (1751); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) inhibited pacing.Additionally, farfield r wave oversensing was noted.Technical services (ts) was consulted to review an atrial tachy response (atr) episode and found that atrial pacing was inhibited due to atrial flutter response (afr).Ts found regular farfield r wave oversensing in events and determined it was less persistent when a mode switch occurred but was still present.Ts offered next steps to establish a root cause for the pacing inhibition.No adverse patient effects were reported.This crt-d remains in service.
 
Additional Manufacturer Narrative
This product investigation is still in progress.This report will be updated when product investigation is complete.If pertinent information is provided in the future, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM? X4 CRT-D
Common Device Name
Implantable Cardiac Resynchronization Therapy Defibrillator
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50839789
Report Number2124215-2026-50156
Device Sequence Number10262747
Product Code NIK
UDI-Device Identifier00802526589188
UDI-Public(01)00802526589188(17)260513
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/13/2026
Device Model NumberG138
Device Catalogue NumberG138
Device Lot Number395926
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2026
Supplement Date Received by Manufacturer09/21/2026
Initial Report FDA Received Date09/21/2026
Supplement Report FDA Received Date09/24/2026
Date Device Manufactured05/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
-
-