It was reported that during a laparoscopic lobectomy, the surgeon found that a large amount of the staples fell off and scattered in the thoracic cavity after the device firing was completed; and only a small amount of the staples remained in the marginal tissue of the anastomosis.Immediately and completely counted and removed all scattered staples, and repeatedly flushed the thoracic cavity with normal saline to confirm that there is no residual foreign body.Changed another one to continue the surgery but the same problem happened again.Changed another one to continue the surgery, it was reported that during the surgery, could not fire.Changed the new one to complete surgery.No additional information can be provided.
|
|
(b)(4)
date sent: 9/21/2026
d4: batch #862d78.Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one gst60b reload was received partially fired 1/4 and with an opened package.The returned reload was reset and loaded into a test device.The returned reload was tested for functionality with a test device in the straight position and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples met the staple form release criteria.The partially fired is consistent with an incomplete or interrupted cycle.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 862d78, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|