Ethicon Endo-Surgery, LLC. Echelon Endopath; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number GST60G
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/10/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unspecified surgical procedure, it was observed that the device could not be fired.Changed to a new one to complete the procedure.There was no patient consequence nor surgical delay reported.No additional information provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 9/21/2026.D4: batch #950d35.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one gst60g reload was received fully fired.Upon further inspection, the reload was found to have damaged on the knife channel, it is possible that the knife may push the staple line and tissue forward shaving the reload knife channel.No functional test could be performed due to the condition of the reload.To mitigate the potential for staples getting into the reload and interfering with the knife path during device firing, prior to reloading the device, rinse the anvil and reload jaw in sterile solution and then wipe the anvil and reload jaw to clean any formed but unused staples from the device.Additionally, proper care should be taken when placing the device on the tissue to be stapled, to ensure that no hard obstruction such as a clip is included with the tissue inside the jaws.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 950d35, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number of 969d85 / 25gst60g , and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Echelon Endopath
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR  00969
Manufacturer (Section G)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 00969
PR   00969
+1(303)552-6892
MDR Report Key50840269
Report Number3005075853-2026-07766
Device Sequence Number10039723
Product Code GDW
UDI-Device Identifier10705036014713
UDI-Public(01)10705036014713(17)290331(10)969D85
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K183435
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Type of Report Initial
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberGST60G
Device Lot Number969D85
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/02/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2026
Initial Report FDA Received Date09/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/08/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-