MAUDE Adverse Event Report: Sistemas Medicos Alaris S.A de C.V. BD MaxZero Pressure Rated Extension Set; Intravascular Administration Set

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Sistemas Medicos Alaris S.A de C.V. BD MaxZero Pressure Rated Extension Set; Intravascular Administration Set Back to Search Results
Catalog Number MZ5303
Medical Device Problem Codes Fluid/Blood Leak (1250); Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the bd maxzero pressure rated extension set had connection issues.The following information was provided by the initial reporter: a cathena iv was placed and a maxzero extension set was connected.The connection would not seal properly and blood leaked at the connection until the extension set was replaced.Extension set product information: material number: mz5303.Material description: bd maxzero.Batch number: 26045881.Additional information received from the customer: can you please provide an exact date of event in this format mm-dd-yyyy? date was 08-27-2026.Describe any patient harm, injury, complication or negative outcome that occurred as a result of the event.Minor blood loss (only a few ml), delay to blood sample collection and saline infusion, extension set had to be replaced with another one from the same lot.
 
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MaxZero Pressure Rated Extension Set
Common Device Name
Intravascular Administration Set
Manufacturer (Section D)
Sistemas Medicos Alaris S.A de C.V.
parque sur, el realito
blvd. 2000
tijuana 22250
MX  22250
Manufacturer (Section G)
Sistemas Medicos Alaris S.A de C.V.
parque sur, el realito
blvd. 2000
tijuana 22250
MX   22250
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
+1(847)393-5694
MDR Report Key50840322
Report Number9616066-2026-02381
Device Sequence Number16018183
Product Code FPA
UDI-Device Identifier10885403230561
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K140831
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberMZ5303
Device Lot Number26045881
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2026
Initial Report FDA Received Date09/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/29/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-