(b)(4).Date sent: 9/21/2026.D4 batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, and the following was obtained:
did the device deliver any staples?
yes.If yes, were the staples formed properly?
yes, but only 1/4th of the way down the staple load.If yes, was the staple line complete?
no.Did the device cut?
no.Was there any difficulty opening the device?
yes, the emergency release failed.If yes, how was the device removed from the patient?
the second attempt of the emergency release worked
if yes, did the jaws eventually open?
yes.Did any component fell into the patient?
no.Did retrieval alter the procedure in any way?
yes.If yes, please explain.This caused the procedure to go longer than necessary.Is the current patient status known?
yes, no harm was caused to the patient.Was there any patient consequence or change in the post-operative care of the patient as a result of the event?
no.
|
|
(b)(4).Date sent: 10/07/2026
d4: batch #951d76.Correction: h4.Investigation summary:
the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one psee45a device was returned with no apparent damage and with a gst45w reload present.The reload was received partially fired 1/10.The returned device and reload were tested for functionality in the straight position.The reload was reset and reloading it into the device.The device achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples met the staple form release criteria.It is possible that while loading the reload, the reload was pushed farther back than the reload alignment slot resulting in the knife pushing the one-piece sled forward and locking the reload.The event described of difficult to open and difficult to actuate manual override could not be confirmed as the device performed and opened as expected.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 951d76, and no non-conformances were identified.
|