Catalog Number ECR60D
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unknown procedure, it was observed that the device could not be fired.Changed another one to continue the surgery, the same problem happened again.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 9/21/2026.D4: batch #161c07.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ecr60d reload was received fully fired, with the pan detached from the reload.No functional test was performed due to the condition of the reload.It is also possible that handling by the customer could dislodge the pan.Dislodgement as a result of the removal of the reload from the device is the most likely scenario.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number of 314c50/22ecr60d , and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 161c07, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON; ENDOPATH
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR  00969
Manufacturer (Section G)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 00969
PR   00969
+1(303)552-6892
MDR Report Key50840879
Report Number3005075853-2026-07776
Device Sequence Number10270694
Product Code GDW
UDI-Device Identifier10705036011842
UDI-Public(01)10705036011842
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K070887
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Type of Report Initial
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Catalogue NumberECR60D
Device Lot Number314C50
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/02/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2026
Initial Report FDA Received Date09/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1