The complaint stated, "the clinicians and staff have reported that we are having problems holding pressure in one of our products we get from medline." medline quality received one unit for sample evaluation; the bag was inspected for any defects and none were found, the bag was filled to 300mmhg and closed off with the attached clamp, the bag was compressed and displayed no signs of pressure loss, and the clamp was moved to the "deflate" position and the bag deflated with no issue; this cycle was repeated two more times with the same result.Medline was not able to confirm the quality issue about the pressure infuser bag not holding pressure, and the cause was documented as not confirmed with received sample.The overall conclusion was that the quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|