| Model Number |
8637-20 |
| Medical Device Problem Codes |
Communication or Transmission Problem (2896); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received from a patient who was receiving clonidine, bupivacaine, and dilaudid (hydromorphone) via an implantable pump for non-malignant pain indications for use.It was reported that the communicator port and charging cord had melted.Patient stated the plastic on both sides was damaged, and the connection appeared burnt and dark.Patient added they were about to perform a bolus but noticed the communicator and charging cord had melted.Agent sent a request to repair to replace the communicator and charging cord.Additional information was received when the patient received their replacement communicator.The patient noted that there were no instructions included. agent attempted to walk patient through pairing communicator.Patient reported not found communicator and tapped switch communicator.Patient reported searching, ensure the communicator is powered on and then no device found with blue screen and the option to retry.Patient confirmed communicator was powered on and not plugged in and toggled bluetooth off and back on; patient confirmed location was on.Patient again attempted to pair and reported same screens.Patient restarted handset and reset communicator and again reported same screens.Agent warm transferred to health care it support services.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d information references the main component of the system.Other relevant device(s) are: product id: tm90t0, serial/lot #: (b)(6), ubd: (b)(6)2022, udi#: (b)(4) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
H3:tm90t0 - npa134200n: the tm90t0 communicator was analyzed on (b)(6)2026.Visual observation showed micro-usb charge port was loose.The plastic near micro usb port was melted.The device failed battery charging bench test and final function jbal test.The device was scrapped and taken out of service medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|