| Model Number |
1001780 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the procedure was to treat a lesion in the right coronary artery.The hi-torque bmw guide wire was advanced into the patient however the radiopaque tip was not visible on fluoro as it had separated.As a result, the procedure could not be completed due to the absence of the radiopaque feature.There were no adverse patient effects.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.As there was no damage noted to the device during the inspection prior to use, it is possible that inadvertent interaction with the anatomy/other devices and/or inadvertent mishandling resulted in the reported tip material separation; however this cannot be confirmed.The investigation determined a conclusive cause for the reported tip material separation cannot be determined.Based on the reported information and results of the complaint investigation there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|