| Model Number |
Model 200 |
| Medical Device Problem Code |
Visual Prompts will not Clear (2281)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the autopulse nxt platform (sn 01808) showed an alert yellow triangle led during the resuscitation attempt.The performance report showed alert 120 (analog motor temperature alert) and fault code 1290 (analog motor temperature over the thermal limit).No delay in care was caused; the crew switched to manual cpr when the autopulse nxt platform stopped working and continued the transport.No consequences or impacts on the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
For g4: pma/510(k): premarket submission number not available/not released.The autopulse nxt platform involved in the reported complaint will not be returned for evaluation because the customer conducted a very long bench test on the nxt platform using a manikin and noted no errors or unusual heat on the exhaust side.At this point, the customer assumes something interfered with airflow on the platform during the reported event.The platform has been returned to service for clinical use, and the customer will continue to monitor.The customer has provided the archive data from the autopulse nxt platform.Zoll is reviewing the archive data.A follow-up report will be submitted once the archive data review is complete.
|
| |
|